Ultra Stable Molecular Sensors by Submicron Referencing and Why They Should Be Interrogated by Optical Diffraction—Part II. Experimental Demonstration
Part II: an unstabilized diffractometric biosensor compared directly with a Biacore 8k instrument
- Frutiger et al.
- Sensors
- Dec 2020
Frutiger et al. show experimentally that sensors using the spatial affinity lock-in and diffraction readout overcome drawbacks of refractometric biosensors. They compare an unstabilized diffractometric biosensor directly with a Biacore 8k instrument.
The Challenge: Cross-Sensitivity of Refractometric Biosensors
Surface plasmon resonance is the standard in label-free interaction analysis, despite a significant cross-reactivity to environmental and experimental noise. Part I proposed the spatial affinity lock-in as the answer. This paper tests the idea in the lab.
The Approach: Direct Comparison and a Combined Sensor
The authors compared the sensitivity to refractive index jumps and the surface mass resolution of both sensor types. A combined refractometric and diffractometric biosensor then measured both signals in the same experiments. A conceptual and quantitative discussion explains the physical reasons and defines a figure of merit.
Key Results Against Refractometric Detection
- Refractive index jumps: a 0.03 RIU discontinuity lowered the diffractometric signal by 2 to 3 pg/mm². That is more than three orders of magnitude below the refractometric response in the same experiment.
- Cross-sensitivity: calculated at 69.3 pg/mm²/RIU in water for the diffractometric sensor, against a measured 760 pg/mm²/RIU for the referenced Biacore 8k output.
- Drift: below 100 fg/mm² for the unstabilized diffractometric sensor, with some of the best Biacore flow cells on par.
- Combined sensor: baseline noise of 0.1 pg/mm² rms diffractometric against 100 pg/mm² rms refractometric.
- Precision: for the same mass resolution, a refractometric sensor needs a relative precision up to four orders of magnitude higher.
Why It Matters for Label-Free Sensing
Low-precision, unstabilized diffractometric devices reached the resolution of large stabilized refractometric sensors. The authors believe label-free optical sensors could therefore move beyond the drug discovery lab. They point to miniaturized, mass-produced environmental and medical sensors as possible uses.
More from our resource library you may like
Oct 2026
High-throughput thermodynamic fingerprinting of protein–ligand interactions by DNA-directed focal molography
Oehninger, Notova, Frutiger
Oehninger, J., Notova, S., & Frutiger, A. (2026). PLOS One, 21(10), e0354008. https://doi.org/10.1371/journal.pone.0354008
Download after a short form
Oct 2026
Same target, different stickiness: polyreactivity of therapeutic antibodies
lino Biotech AG
Six clinical antibodies, including two same-target pairs, ranked by polyreactivity against an eight-ligand panel on one focal molography chip. Briakinumab scored 69-fold above ...
Download after a short form
Oct 2026
One on-DNA ligand against many targets: selectivity profiling for DEL hit validation
lino Biotech AG
On-DNA oncoFAP profiled against human, mouse and canine FAP and five unrelated proteins on one focal molography chip layout. Full kinetics on every species variant, flat responses ...
Download after a short form