Ultra Stable Molecular Sensors by Submicron Referencing and Why They Should Be Interrogated by Optical Diffraction—Part II. Experimental Demonstration

Part II: an unstabilized diffractometric biosensor compared directly with a Biacore 8k instrument

  • Frutiger et al.
  • Sensors
  • Dec 2020
Get the publication

Frutiger et al. show experimentally that sensors using the spatial affinity lock-in and diffraction readout overcome drawbacks of refractometric biosensors. They compare an unstabilized diffractometric biosensor directly with a Biacore 8k instrument.

The Challenge: Cross-Sensitivity of Refractometric Biosensors

Surface plasmon resonance is the standard in label-free interaction analysis, despite a significant cross-reactivity to environmental and experimental noise. Part I proposed the spatial affinity lock-in as the answer. This paper tests the idea in the lab.

The Approach: Direct Comparison and a Combined Sensor

The authors compared the sensitivity to refractive index jumps and the surface mass resolution of both sensor types. A combined refractometric and diffractometric biosensor then measured both signals in the same experiments. A conceptual and quantitative discussion explains the physical reasons and defines a figure of merit.

Key Results Against Refractometric Detection

  • Refractive index jumps: a 0.03 RIU discontinuity lowered the diffractometric signal by 2 to 3 pg/mm². That is more than three orders of magnitude below the refractometric response in the same experiment.
  • Cross-sensitivity: calculated at 69.3 pg/mm²/RIU in water for the diffractometric sensor, against a measured 760 pg/mm²/RIU for the referenced Biacore 8k output.
  • Drift: below 100 fg/mm² for the unstabilized diffractometric sensor, with some of the best Biacore flow cells on par.
  • Combined sensor: baseline noise of 0.1 pg/mm² rms diffractometric against 100 pg/mm² rms refractometric.
  • Precision: for the same mass resolution, a refractometric sensor needs a relative precision up to four orders of magnitude higher.

Why It Matters for Label-Free Sensing

Low-precision, unstabilized diffractometric devices reached the resolution of large stabilized refractometric sensors. The authors believe label-free optical sensors could therefore move beyond the drug discovery lab. They point to miniaturized, mass-produced environmental and medical sensors as possible uses.